B • Installation
Doc. 065125-01 10/06
77
B.3.12 Connecting the Internal Peristaltic Pump (Optional)
The internal peristaltic pump (i.e., the AS-HV rinse pump) can be used
for sample loading if the rinse station is connected directly to an external
deionized water source.
NOTE When using an external rinse source, the delivery
pressure should be less than 0.03 MPa (5 psi).
1. Connect the peristaltic tubing from the rinse reservoir to a deionized
water source.
2. Connect the liquid lines to the pump as required for the push mode
3. When you finish the plumbing connections, go on to
.
B.3.13 Connecting a Relay-Controlled Pump (Optional)
An external relay-controlled pump from a third-party vendor can be used
for sample loading. Refer to the pump manual for detailed installation
instructions.
1. Place the relay-controlled pump near the AS-HV on the lab bench.
2. Connect the relay cable (P/N 064350) from the pump to Relay B on
the AS-HV side panel.
3. When you finish the plumbing connections, go on to
.