3 • Operation and Maintenance
Doc. 065125-01 10/06
37
3.3
Routine Maintenance
This section describes routine maintenance procedures that the user can perform
for the AS-HV and the syringe pump. Any maintenance procedures for the AS-
HV and the syringe pump that are not described here must be performed by
Dionex personnel. For maintenance procedures for other sample loading pumps,
refer to the respective product manual.
To contact Dionex in the U.S., call 1-800-346-6390 and select the Technical
Support option. Outside the U.S., call the nearest Dionex office.
NOTE The AS-HV has no lubrication requirements. Do not
apply lubrication to the sampling arm or any other
components.
3.3.1
Daily Maintenance
•
Wipe up all spills immediately, especially spills of corrosive liquids.
•
Inspect the syringe pump for leaks, and correct any problems.
•
Flush the syringe pump thoroughly with distilled or deionized water
after each use and when the pump is not in use.
NOTE Do not allow the syringe pump to run dry for
more than a few cycles.
•
Check for air bubbles in the syringe(s) and tubing and remove any
bubbles that appear (see
•
Replace the rinse fluid. Dionex recommends using ASTM Type I (or
better) filtered, deionized water as the rinse fluid.
•
To ensure correct drainage, the end of the waste line should not be
submerged in waste liquid. Check the volume of liquid in the waste
container and empty when needed.
Neutralize acidic and caustic wastes before disposal. Dispose of all
wastes in accordance with local regulations.
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...