AS-HV Operator’s Manual
48
Doc. 065125-01 10/06
5.2
Replacing the Sample Syringe
Follow the instructions here to install a new 10 mL syringe (P/N 064224) in the
syringe pump.
5.2.1
Removing the Existing Syringe
Refer to
when removing the existing syringe.
1. Remove the liquid from the
syringe.
2. Remove the plunger lock
screw.
3. In Chromeleon, select
SyringeAspirate
in the
Commands dialog box.
Select the maximum
Volume
(i.e., the size of the installed
syringe) and click
OK
to
lower the plunger drive.
4. Unscrew the syringe from the
valve.
Figure 5-1. Removing the Syringe
Remove the plunger lock
screw
Valve
Syringe