3 • Operation and Maintenance
Doc. 065125-01 10/06
29
Table 3-3. “Slow” Syringe Speed Settings: Wash Port Dispense
Speed for Sample Loop/Push or Pull Mode
Speed Setting
Top Speed
(steps/sec)
Flow Rate
(mL/min)
1
80
1.00
2
160
2.00
3
200
2.50
4
240
3.00
5
280
3.50
6
320
4.00
7
360
4.50
8
400
5.00
9
440
5.50
10
480
6.00
Table 3-4. “Slow” Syringe Speed Settings: Sample Aspirate Speed
for Sample Loop/Push or Pull Mode or Concentrator Column/Push
Mode
Speed Setting
Top Speed
(steps/sec)
Flow Rate
(mL/min)
1
32
0.40
2
64
0.80
3
96
1.20
4
128
1.60
5
160
2.00
6
192
2.40
7
224
2.80
8
256
3.20
9
288
3.60
10
320
4.00