C • Error Messages and Codes
Doc. 065125-01 10/06
99
Cause:
The
PrepareNextSample
command cannot be executed until after the sample
has been injected.
To troubleshoot:
Edit the Chromeleon program so that
PrepareNextSample
appears after the
Inject
command.
Cause:
You selected a sample position number that is invalid.
To troubleshoot:
Enter a valid position number. The AS-HV supports both a 15-position
sample rack and a 24-position sample rack. Valid position numbers for the
racks are R1 to R15 or R1 to R24, respectively.
Cause:
The syringe pump is receiving new Chromeleon commands before a previous
command has been executed.
To troubleshoot:
Allow enough time for each command in the program to be executed before
the next command starts.
PrepareNextSample may only be specified after the Inject
command.
Sample number error.
Syringe pump command overflow.