5 • Service
Doc. 065125-01 10/06
51
5.2.4
Initializing the New Syringe
1. Turn off the AS-HV power,
wait 10 seconds, and turn on the
power again.
2. In Chromeleon, select
SyringeReset
in the Commands
dialog box and click
OK
.
This moves the syringe plunger
down slightly, into its initial
position (see
Figure 5-3. Initializing the New Syringe
Initial
Position