
5 • Service
Doc. 065125-01 10/06
51
5.2.4
Initializing the New Syringe
1. Turn off the AS-HV power,
wait 10 seconds, and turn on the
power again.
2. In Chromeleon, select
SyringeReset
in the Commands
dialog box and click
OK
.
This moves the syringe plunger
down slightly, into its initial
position (see
Figure 5-3. Initializing the New Syringe
Initial
Position
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...