C • Error Messages and Codes
Doc. 065125-01 10/06
97
Cause:
You created a Chromeleon program that does not include pressure limit
settings for the AXP pump.
To troubleshoot:
Always specify both an upper and lower pressure limit for the AXP in your
programs.
Cause:
You entered a parameter that is outside the range allowed by Chromeleon.
To troubleshoot:
Enter a valid setting for this parameter.
Cause:
The
Volume
entered for
SyringeAspirate
or
SyringeDispense
in the
Commands dialog box exceeds the amount of sample in the syringe.
To troubleshoot:
Check the sample volume in the syringe, and then enter the correct volume.
Missing AXP pump pressure limits in the PGM.
Parameter out of range.
Parameter out of range; check syringe position.