C • Error Messages and Codes
Doc. 065125-01 10/06
93
and then wait until the
CPU
LED flashes. If this is successful, try
again to send commands from Chromeleon to the AS-HV. If the
HOST
LED still fails to flash, contact Dionex for assistance.
Cause:
A problem with the AS-HV firmware.
To troubleshoot:
The AS-HV firmware cannot be serviced by the user. Contact Dionex for
assistance.
Cause:
A problem with the AS-HV firmware or electronics resulted in the sampling
arm being commanded to move to a position outside of its operating envelope.
To troubleshoot:
Initialize the AS-HV. (Turn off the power, wait 10 seconds, and turn on the
power again.) If the message appears again, contact Dionex for assistance.
Cause:
The AS-HV firmware received an unrecognized command.
To troubleshoot:
1. Verify that Chromeleon release 6.80 (or later) is installed; software
releases prior to this do not support the AS-HV.
2. Check that the cable between the AS-HV and the Chromeleon PC is
firmly connected and screwed into place.
Configuration value out of range.
Envelope error.
Invalid command.