2 • Description
Doc. 065125-01 10/06
9
I/O Port
The I/O port provides eight contact closure inputs and four relay
outputs.
The autosampler has four internal relays (A, B, C, and D).
Normally open and normally closed contact pairs are available on
the external I/O connector.
Relay A is reserved for control of the internal peristaltic pump.
Relay B is reserved for control of an external relay-controlled
pump operating as the sample loading pump. The relays can be
controlled through commands programmed in Chromeleon.
RELAY Status Indicators
These indicators light (yellow) to indicate the
state of the internal relays (A, B, C, and D).
When a relay is energized, the corresponding
indicator is lit.
The relay outputs can be used to connect and control low power devices. These
outputs should be limited to loads of 1 amp or less. The eight digital inputs
available on the autosampler allow monitoring of electrical signals between 0 and
+24 volts.
System
Status Indicators
These indicators light (yellow) to identify certain
autosampler conditions. For details, see the table below.
.
Status Indicator
Description
Indication
S1
Programmable behavior
LED
ON when programmed
AUX
Auxiliary instrument port
activity LED
ON when a byte is sent or received
on Auxiliary Instrument Port, then
OFF
HOST
Host interface port
activity LED
ON when a byte is sent or received
on Host Interface Port, then OFF
CPU
Central processor unit
(CPU) activity LED
OFF for 10 seconds, then ON for
1 second
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...