AS-HV Operator’s Manual
16
Doc. 065125-01 10/06
2.4.4
Internal Peristaltic Pump
The AS-HV rinse station includes a built-in peristaltic pump that provides
a supply of fresh rinse fluid to the reservoir (see
controlled through Relay A.
If the rinse station is connected directly to an external rinse source, the
peristaltic pump can be replumbed for use as the sample loading pump in
either the push or pull mode (see
instructions, refer to
To ensure optimal performance when using the internal peristaltic pump
for sample loading, the delivery pressure should be less than 0.03 MPa
(5 psi).
2.4.5
Relay-Controlled Pump
An external relay-controlled pump from a third-party vendor can be used
for sample loading in either the push or pull mode (see
pump is controlled through Relay B.
A relay cable connects the pump to the
I/O
port on the AS-HV side panel
). Note that the pump is supported by Chromeleon,
although only through relay control.
If you have a question regarding the use of third-party pumps with the
AS-HV, please contact Dionex. In the U.S., call 1-800-346-6390 and
select the Technical Support option. Outside the U.S., call the nearest
Dionex office.
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...