AS-HV Operator’s Manual
88
Doc. 065125-01 10/06
Cause:
If a Chromeleon program includes the
PrepareNextSample
command, it
must be the last AS-HV command in the program.
To troubleshoot:
Delete any AS-HV commands that appear after
PrepareNextSample
.
Cause:
A problem with the AS-HV firmware.
To troubleshoot:
Initialize the AS-HV. (Turn off the power, wait 10 seconds, and turn on the
power again.) If the message appears again, the AS-HV firmware and/or
variables must be reloaded. Contact Dionex for assistance.
After PrepareNextSample, no further sampler commands are
supported.
AS-HV configuration memory checksum error.
AS-HV program memory checksum error.