5 • Service
Doc. 065125-01 10/06
57
3. Use a thin blade screwdriver to unclip the
cover of the fuse compartment (see
Figure 5-8. Opening the Fuse Compartment
4. Slide the fuse holder out of the input panel
Figure 5-9. Removing the Fuse Holder
5. Remove the fuses from the fuse holder and replace with new 5 amp slow-
blow fuses (P/N 064248). Dionex recommends always replacing both fuses.
6. Slide the fuse compartment back into the unit and clip the cover into place.
7. Reconnect the power cord and turn on the power.
Insert
screwdriver to
unclip cover
Fuse