Doc. 065125-01 10/06
85
C • Error Messages and Codes
•
describes error messages that may be reported by Chromeleon
software and provides guidelines for troubleshooting problems.
•
describes error codes that may be displayed by the AS-HV
ERROR
indicators and provides guidelines for troubleshooting problems.
NOTE All firmware upgrades and repairs of electronics
components must be performed by Dionex personnel.
C.1
Audit Trail Messages
The AS-HV firmware periodically checks the status of certain operating
parameters. If a problem is detected, it is reported to Chromeleon and displayed in
the Audit Trail.
Each error message is preceded by an icon that identifies the seriousness of the
underlying problem (see the table below). You can change the severity level
assigned to a problem whenever appropriate.
Icon
Severity Level
Description
Warning
A message is displayed in the Audit Trail, but the
current run is not interrupted.
Error
A message is displayed in the Audit Trail and the
system attempts to correct the problem (sometimes
by using an alternative parameter). An Error never
interrupts the current analysis; however, if it occurs
during the Ready Check, the analysis will not be
started.
Abort
A message is displayed in the Audit Trail and the
running batch is aborted.
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...