3 • Operation and Maintenance
Doc. 065125-01 10/06
31
To enable control of the injection valve by another device in the
timebase, select the
Valve Control
check box. When the check box is
selected, the following options are enabled:
•
Load/Inject Valve Using:
Select the injection valve or relay/TTL
device that will control sample loading and injection.
•
Injection Duration:
Specify for how long injection occurs. (This
setting is ignored unless you select
Sample Loop
as the sample
loading type in the Server Configuration (see
NOTE For sample overlap programs, the Injection
Duration must be less than the Delay Before
Processing Next Sample (see below).
To enable sample overlap (see
), select the
Sample
Overlap
check box. When the check box is selected, the following
option is enabled:
•
Delay Before Processing Next Sample:
Specify for how long
the Chromeleon program waits in order to finish the currently
running sample before executing the remaining commands in the
program.
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...