AS-HV Operator’s Manual
66
Doc. 065125-01 10/06
4. Reposition the track over the rotor and fix by slotting the track
occlusion pin back into the track slot.
NOTE Compressing the spring on the track occlusion
pin will increase the suction height and
performance against backpressure.
5. Connect the delivery line to the inlet nipple of the fixed rinse
). Keep the delivery line as short as possible.
6. Connect the suction line to the reservoir containing rinse fluid. Keep
the suction line as short as possible.
NOTE Dionex recommends using ASTM Type I (or
better) filtered, deionized water as the rinse
fluid.
To remove tubing from the rinse pump:
•
Release the track by disengaging the track occlusion arm.
•
Remove both tubing bridges from their location slots and remove the
tubing from the pump head.
B.3.5
Installing the Standards Rack
The AS-HV is shipped with an 11-position standards rack (P/N 064253)
(see
). An optional 6-position standards rack (P/N 064252) is
available, also.
1. Check the rack location mat
for the location of the three
mounting holes that are closest
to the AS-HV chassis. (The
holes may be positioned at the
rear and/or the front of the
mat.)
Figure B-5. Standards Rack (11-position)
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...