AS-HV Operator’s Manual
96
Doc. 065125-01 10/06
Cause:
You selected an invalid position number for the sample rack or standards rack.
To troubleshoot:
Enter a valid position number:
•
The AS-HV supports a 6-position standards rack and an 11-position
standards rack. Valid position numbers for the racks are S1 to S6 or
S1 to S11, respectively.
•
The AS-HV supports a 15-position sample rack and a 24-position
sample rack. Valid position numbers for the racks are R1 to R15 or R1
to R24, respectively.
Cause:
You entered an invalid injection volume.
To troubleshoot:
Enter a value from 0.01 to 250 mL.
Cause:
You created a Chromeleon program in which one of the pressure limit settings
required for the AXP pump is missing.
To troubleshoot:
Always specify both an upper and lower pressure limit for the AXP in your
programs.
Invalid position.
Invalid volume.
Missing AXP pump pressure high limit in the PGM.
Missing AXP pump pressure low limit in the PGM.