3 • Operation and Maintenance
Doc. 065125-01 10/06
35
10. Specify the
State
or
Duration
for the selected device:
•
Clicking
State
specifies a simple switch command. Select
Open
or
Closed
from the drop-down list box.
•
Clicking a state description sets the device to the selected state
and lets you specify a
Duration
. Enter the duration of the
selected state in the edit box.
NOTE Relays A and B are reserved for internal use
and cannot be selected.
11. To add the following information to the list box, click
Add
:
•
Retention Time
value
•
Name of the selected
Device
•
State
or
Duration
value
Chromeleon will automatically organize all entries in the list box in
chronological order.
To remove an entry from the list box, select the entry and click
Delete
.
(All entries in the list box will automatically be reorganized.)
12. Click
Next
to continue.