AS-HV Operator’s Manual
98
Doc. 065125-01 10/06
Cause:
You failed to enter one of more of the required parameters for the
SyringeAspirate
or
SyringeDispense
command.
To troubleshoot:
To execute the
SyringeAspirate
command, you must specify both the source
of the sample to be aspirated and the volume of the syringe installed in the
syringe pump.
To execute the
SyringeDispense
command, you must specify both the
destination for the sample to be dispensed and the volume of the syringe
installed in the syringe pump.
Cause:
The Chromeleon program contains more than one
PrepareNextSample
command for the AS-HV.
To troubleshoot:
Delete all but one
PrepareNextSample
command from the program.
Cause:
You selected the
PrepareNextSample
command in the Chromeleon
Commands dialog box.
To troubleshoot:
Always use the Program Wizard to create sample overlap programs.
Please enter all parameters.
PrepareNextSample can only be issued once per program.
PrepareNextSample cannot be used in a manual program.