AS-HV Operator’s Manual
72
Doc. 065125-01 10/06
Figure B-9. Syringe Pump Flow Schematic: Push Mode
Figure B-10. Syringe Pump Flow Schematic: Pull Mode
1
2
3
4
6
5
Eluent In
To Separator
Column
Sample Loop (or Concentrator Column)
Sample
To Flush
Bottle
To Waste
1.0-mm (0.040-in) ID
PEEK Tubing
Syringe
Pump
Sample Probe
Injection
Valve
A
A
1/4-28 to 10-32 coupler, double-cone ferrule fitting, and
10-32 double-cone bolt
1.57-mm (0.062-in) ID
PFA Tubing
1.57-mm (0.062-in) ID
PFA Tubing
1
2
3
4
6
5
To Separator
Column
Eluent In
Sample Loop (or Concentrator Column)
Sample
To Flush
Bottle
To Waste
1.0-mm (0.040-in) ID
PEEK Tubing
Syringe
Pump
Sample Probe
Injection
Valve
A
A
1.57-mm (0.062-in) ID
PFA Tubing
1.57-mm (0.062-in) ID
PFA Tubing
1/4-28 to 10-32 coupler, double-cone ferrule fitting, and
10-32 double-cone bolt