Actions:
•
Check for a crack in the syringe. If the syringe is cracked, it must be
replaced (see
•
Check the seal for scratches, deformation, or other damage. If the seal
is damaged, the syringe must be replaced (see
•
If the syringe is not cracked and the seal is
not damaged, check that the syringe is
tightened. To tighten, hold the syringe at
the top and bottom fittings and turn it
clockwise (as viewed from the top) (see
). Tighten fingertight only; do
not overtighten.
Figure 4-1. Tightening the Syringe
4.1.3
Leaking Waste Line (Fixed Rinse Reservoir)
•
Check that the waste line is pushed tightly onto the drip tray fitting
and that it is not elevated at any point after it exits the AS-HV.
•
Check that the waste line is not crimped or otherwise blocked.
•
To ensure correct drainage, make sure the end of the waste line is not
submerged in waste liquid.
•
Check the volume of liquid in the waste container; empty as needed.
Neutralize acidic and caustic wastes before disposal. Dispose of all
wastes in accordance with local regulations.