AS-HV Operator’s Manual
44
Doc. 065125-01 10/06
4.6
Power Supply Fails
If the green power indicator on the front of the AS-HV and/or the system status
indicator LEDs on the AS-HV side panel fail to light when expected, it may
indicate that the AS-HV power supply is not operational.
Corrective action:
1. Check that the power cable is connected to the AS-HV power entry and that
the power switch is turned on.
2. Check the fuses. If a fuse is blown, replace
both
fuses with new 5 amp slow-
blow fuses (P/N 064248). For fuse replacement instructions, refer to
.
3. If the autosampler still fails to power up, contact Dionex for assistance.
NOTE Access to the power supply unit and components
requires removal of the main housing. Do not attempt to
access these components. All repairs of the electronics
must be performed by Dionex personnel.
For continued protection against risk of fire or shock, replacement
fuses must be the type and rating specified here.
Pour maintenir la protection contre les risques d'incendie ou
d'électrocution, remplacez toujours les fusibles par des fusibles du
même type et du même calibre.
Zum Schutz vor Feuer und Stromschlägen müssen beim
Sicherungswechsel immer Sicherungen des gleichen Typs und mit
gleicher Leistung verwendet werden.
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...