450
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
24
Mask Test Commands
:MTESt:FOLDing
(Clock Recovery software only)
Command
:MTESt:FOLDing {{ON | 1} | {OFF | 0}}
The :MTESt:FOLDing command enables (ON) or disables (OFF) the display
of the real time eye. When enabled, an eye diagram of the data will be
generated syperimposed on the recovered clock. Refer to the
:MEASure:CLOCk commands for clock recovery.
Example
This example enables the display of the real time eye.
myScope.WriteString ":MTEST:FOLDING ON"
Query
:MTESt:FOLDing?
The :MTESt:FOLDing? query returns the current state of clock recovery
folding.
Returned Format
[:MTESt:FOLDing] {1 | 0} <NL>
Example
myScope.WriteString ":MTEST:FOLDING?"
varValue = myScope.ReadNumber
Debug.Print FormatNumber(varValue, 0)
See Also
•
":MEASure:CLOCk"
on page 495
•
":MEASure:CLOCk:METHod"
on page 496
•
":MEASure:CLOCk:METHod:ALIGn"
on page 498
•
":MEASure:CLOCk:METHod:DEEMphasis"
on page 499
•
":MEASure:CLOCk:METHod:EDGE"
on page 500
•
":MEASure:CLOCk:METHod:JTF"
on page 502
•
":MEASure:CLOCk:METHod:OJTF"
on page 504
N O T E
This command is only available when the E2688A Clock Recovery Software is installed.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...