158
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
9
Acquire Commands
This number determines how many samples are averaged together to form
the 16- bit samples that are stored into the channel memories.
To set the desired bits of vertical resolution, see
":ACQuire:HRESolution"
on page 155.
SEGMented
In this sampling mode you can view waveform events that are separated
by long periods of time without capturing waveform events that are not of
interest to you.
SEGPdetect
Enables Peak Detect Segmented mode.
SEGHres
Enables High Resolution Segmented mode.
To set the desired bits of vertical resolution, see
":ACQuire:HRESolution"
on page 155.
Example
This example sets the acquisition mode to Real Time Normal.
myScope.WriteString ":ACQuire:MODE RTIMe"
Query
:ACQuire:MODE?
The :ACQuire:MODE? query returns the current acquisition sampling mode.
Returned Format
[:ACQuire:MODE] {RTIMe | PDETect | HRESolution | SEGMented}<NL>
Example
This example places the current acquisition mode in the string variable,
strMode, then prints the contents of the variable to the computer's screen.
Dim strMode As String
' Dimension variable.
myScope.WriteString ":ACQuire:MODE?"
strMode = myScope.ReadString
Debug.Print strMode
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...