![Agilent Technologies Infiniium 90000 Q-Series Programmer'S Manual Download Page 837](http://html1.mh-extra.com/html/agilent-technologies/infiniium-90000-q-series/infiniium-90000-q-series_programmers-manual_2868674837.webp)
Trigger Commands
33
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
837
:TRIGger:HOLDoff:MAX
Command
:TRIGger:HOLDoff:MAX <holdoff_time>
This command is only used when you set the :TRIGger:HOLDoff:MODe
command to RANDom. The RANDom mode varies the trigger holdoff from
one acquisition to another by randomizing the time values between
triggers. The randomized values can be between the values specified by the
:TRIGger:HOLDoff:MAX and :TRIGger:HOLDoff:MIN commands.
The Random holdoff mode ensures that the oscilloscope re- arms after each
acquisition in a manner that minimizes or eliminates the likelihood of
triggering at the beginning of a DDR burst. Randomizing the holdoff
increases the likelihood that the oscilloscope will trigger on different data
phases of a multiphase (8 data transfer) burst. This mode mixes up the
traffic pattern the oscilloscope triggers on and is very effective when used
on repeating patterns.
<holdoff_time>
A real number for the maximum random holdoff time.
Query
:TRIGger:HOLDoff:MAX?
The query returns the current maximum holdoff value for the random
holdoff mode.
Returned Format
[:TRIGger:HOLDoff:MAX] <holdoff><NL>
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...