Measure Commands
25
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
581
:MEASure:NOISe:ALL?
Query
:MEASure:NOISe:ALL? {ZERO | ONE}
The :MEASure:NOISe:ALL? query returns the NOISe measurement results
for the specified level. These values are returned as comma separated
values using the following format:
Returned Format
[:MEASure:NOISe:ALL<space>]
TI(<format>),<result>,<state>,
RN(<format>),<result>,<state>,
DI(<format>),<result>,<state>,
PI(<format>),<result>,<state>,
ABUI(<format>),<result>,<state>,
BUI(<format>),<result>,<state>,
ISI(<format>),<result>,<state>,
Count,<number_of_bits>,<state>,
Level,<nominal_level>,<state>,
Eye Height(<format>),<result>,<state>,<NL>
<space>
White space (ASCII 32) character.
<format>
The format value tells you something about how the measurement is made.
For instance, TI(1E- 12) means that the TI measurement was derived using
a bit error rate of 1E- 12. A format of (rms) means the measurement is a
root- mean- square measurement. A format of (dd) means the measurement
uses a dual- Dirac delta model to derive the measurement. A format of
(pp) means the measurement is a peak- to- peak measurement.
<result>
The measured results for the NOISe measurements. A value of 9.37
means that the oscilloscope was unable to make the measurement.
<state>
The measurement result state. See
Table 24
for a list of values and
descriptions of the result state value.
<number_of_
bits>
The number of waveform bits that have been measured.
N O T E
This command is only available when the EZJIT Complete software is installed.
N O T E
Whether some of these values are included or not depends on the setting of
:MEASure:NOISe:METHod and :MEASure:NOISe:REPort.
For example, when :MEASure:NOISe:REPort or :MEASure:NOISe:METHod is SPECtral, the
BUI and ABUI values are not returned, and there are two PI values (one "rms" and one
"dd").
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...