![Agilent Technologies Infiniium 90000 Q-Series Programmer'S Manual Download Page 635](http://html1.mh-extra.com/html/agilent-technologies/infiniium-90000-q-series/infiniium-90000-q-series_programmers-manual_2868674635.webp)
Measure Commands
25
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
635
:MEASure:RJDJ:TJRJDJ?
Query
:MEASure:RJDJ:TJRJDJ?
The :MEASure:RJDJ:TJRJDJ? query returns the Total Jitter (TJ), Random
Jitter (RJ), and the Deterministic Jitter (DJ) measurements. These values
are returned as comma separated values using the following format:
Returned Format
[:MEASure:RJDJ:TJRJDJ] TJ(<tj_format>),<tj_result>,<tj_state>,
RJ(<rj_format>),<rj_result>,rj_state,
DJ(<dj_format>),<dj_result>,<dj_state><NL>
<tj_format>
<rj_format>
<dj_format>
The format value tells you something about how the measurement is made.
For instance, TJ(1E- 12) means that the TJ measurement was derived using
a bit error rate of 1E- 12. A format of (rms) means the measurement is a
root- mean- square measurement. A format of (d- d) means the
measurement uses from a dual- Dirac delta model used to derive the
measurement. A format of (p- p) means the measurement is a peak- to- peak
measurement.
<tj_result>
<rj_result>
<dj_result>
The measured results for the RJDJ measurements. A value of 9.37
means that the oscilloscope was unable to make the measurement.
<tj_state>
<rj_state>
<dj_state>
The measurement result state. See
Table 24
for a list of values and
descriptions of the result state value.
Example
This example places the Total Jitter (TJ), Random Jitter (RJ), and the
Deterministic Jitter (DJ) measurements in the strResults variable and
displays it on the computer's screen.
Dim strResult As String
' Dimension variable.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString ":MEASURE:RJDJ:TJRJDJ?"
strResult = myScope.ReadString
Debug.Print strResult
N O T E
This command is only available when the N5400A/N5401A Software is installed.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...