154
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
9
Acquire Commands
:ACQuire:COMPlete:STATe
Command
:ACQuire:COMPlete:STATe {{ON | 1} | {OFF | 0}}
The :ACQuire:COMPlete:STATe command specifies the state of the
:ACQuire:COMPlete mode. This mode is used to make a tradeoff between
how often equivalent time waveforms are measured, and how much new
data is included in the waveform record when a measurement is made.
This command has no effect when the oscilloscope is in real time mode
because the entire record is filled on every trigger. However, in equivalent
time mode, as few as 0 new data points will be placed in the waveform
record as the result of any given trigger event. You set the acquire mode of
the oscilloscope by using the :ACQuire:MODE command.
ON
Turns the COMPlete mode on. Then you can specify the completion
percent.
OFF
When off, the oscilloscope makes measurements on waveforms after each
acquisition cycle, regardless of how complete they are. The waveform
record is not cleared after each measurement. Instead, previous data
points will be replaced by new samples as they are acquired.
Query
:ACQuire:COMPlete:STATe?
The :ACQuire:COMPlete:STATe? query returns the state of the
:ACQuire:COMPlete:STATe mode.
N O T E
The :ACQuire:COMPlete:STATe command is used only when the oscilloscope is operating in
equivalent time mode and a digitize operation is not being performed. The :DIGitize
command temporarily overrides the setting of this mode and forces it to ON.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...