846
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
33
Trigger Commands
:TRIGger:SWEep
Command
:TRIGger:SWEep {AUTO | TRIGgered | SINGle}
The :TRIGger:SWEep command selects the oscilloscope sweep mode. New
programs should use :RUN and :SINGle for run control and this command
for AUTO and TRIGgered for sweep control. The SINGle sweep control
should not be used.
AUTO
When you select AUTO, if a trigger event does not occur within a time
determined by the oscilloscope settings, the oscilloscope automatically
forces a trigger which causes the oscilloscope to sweep. If the frequency of
your waveform is 50 Hz or less, you should not use the AUTO sweep mode
because it is possible that the oscilloscope will automatically trigger before
your waveform trigger occurs.
TRIGgered
When you select TRIGgered, if no trigger occurs, the oscilloscope will not
sweep, and the previously acquired data will remain on the screen.
SINGle
When you select SINGle, if no trigger occurs, the oscilloscope will not
sweep, and the previously acquired data will remain on the screen. Do not
use in new programs.
Query
:TRIGger:SWEep?
The query returns the specified channel's trigger level.
Returned Format
[:TRIGger:SWEep] {AUTO | TRIGgered}<NL>
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...