966
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
34
Waveform Commands
<X increment>
The X increment is the duration between data points on the X axis. For
time domain waveforms, this is the time between points. If the value is
zero then no data has been acquired. (See the
":WAVeform:XINCrement?"
on page 980 query.)
<X origin>
The X origin is the X- axis value of the first data point in the data record.
For time domain waveforms, it is the time of the first point. This value is
treated as a double precision 64- bit floating point number. If the value is
zero then no data has been acquired. (See the
":WAVeform:XORigin?"
on
page 981 query.)
<X reference>
The X reference is the data point associated with the X origin. It is at this
data point that the X origin is defined. In this oscilloscope, the value is
always zero. (See the
":WAVeform:XREFerence?"
on page 983 query.)
<Y increment>
The Y increment is the duration between Y- axis levels. For voltage
waveforms, it is the voltage corresponding to one level. If the value is zero
then no data has been acquired. (See the
":WAVeform:YINCrement?"
on
page 986 query.)
<Y origin>
The Y origin is the Y- axis value at level zero. For voltage waveforms, it is
the voltage at level zero. If the value is zero then no data has been
acquired. (See the
":WAVeform:YORigin?"
on page 987 query.)
<Y reference>
The Y reference is the level associated with the Y origin. It is at this level
that the Y origin is defined. In this oscilloscope, this value is always zero.
(See the
":WAVeform:YREFerence?"
on page 989 query.)
<coupling>
0 for AC coupling. 1 for DC coupling. 2 for DCFIFTY coupling. 3 for
LFREJECT coupling.
<X display
range>
The X display range is the X- axis duration of the waveform that is
displayed. For time domain waveforms, it is the duration of time across
the display. If the value is zero then no data has been acquired. (See the
":WAVeform:XRANge?"
on page 982 query.)
<X display
origin>
The X display origin is the X- axis value at the left edge of the display. For
time domain waveforms, it is the time at the start of the display. This
value is treated as a double precision 64- bit floating point number. If the
value is zero then no data has been acquired. (See the
":WAVeform:XDISplay?"
on page 979 query.)
<Y display
range>
The Y display range is the Y- axis duration of the waveform which is
displayed. For voltage waveforms, it is the amount of voltage across the
display. If the value is zero then no data has been acquired. (See the
":WAVeform:YRANge?"
on page 988 query.)
<Y display
origin>
The Y- display origin is the Y- axis value at the center of the display. For
voltage waveforms, it is the voltage at the center of the display. If the
value is zero then no data has been acquired. (See the
":WAVeform:YDISplay?"
on page 985 query.)
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...