Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
9
Acquire Commands
:ACQuire:SRATe[:ANALog]
(Analog Sample Rate)
Command
:ACQuire:SRATe[:ANALog] {AUTO | MAX | <rate>}
The :ACQuire:SRATe[:ANALog] command sets the analog acquisition
sampling rate.
AUTO
The AUTO rate allows the oscilloscope to select a sample rate that best
accommodates the selected memory depth and horizontal scale.
MAX
The MAX rate enables the oscilloscope to select maximum available sample
rate.
<rate>
A real number representing the sample rate. You can send any value, but
the value is rounded to the next fastest sample rate. For a list of available
sample rate values see
"SRATe Sample Rate Tables"
on page 171.
Interaction
between
:ACQuire:SRATe[:
ANALog] and
:ACQuire:POINts[
:ANALog]
If you assign a sample rate value with :ACQuire:SRATe[:ANALog] or a
points value using :ACQuire:POINts[:ANALog] the following interactions
will occur. "Manual" means you are setting a non- AUTO value for SRATe
or POINts.
Example
This example sets the sample rate to 250 MSa/s.
myScope.WriteString ":ACQuire:SRATe:ANALog 250E+6"
Query
:ACQuire:SRATe[:ANALog]?
The :ACQuire:SRATe[:ANALog]? query returns the current analog
acquisition sample rate.
Returned Format
[:ACQuire:SRATe:ANALog] {<rate>}<NL>
Example
This example places the current sample rate in the string variable,
strSample, then prints the contents of the variable to the computer's
screen.
Dim strSample As String
' Dimension variable.
myScope.WriteString ":ACQuire:SRATe:ANALog?"
SRATe
POINts
Result
AUTO
Manual
POINts value takes
precedence (sample rate is
limited)
Manual
AUTO
SRATe value takes precedence
(memory depth is limited)
Manual
Manual
SRATe value takes precedence
(memory depth is limited)
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...