618
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
25
Measure Commands
:MEASure:RJDJ:ALL?
Query
:MEASure:RJDJ:ALL?
The :MEASure:RJDJ:ALL? query returns all of the RJDJ jitter
measurements. These values are returned as comma separated values using
the following format:
Returned Format
[:MEASure:RJDJ:ALL<space>]
TJ(<format>),<result>,<state>,
RJ(<format>),<result>,<state>,
DJ(<format>),<result>,<state>,
PJ(<format>),<result>,<state>,
BUJ(<format>),<result>,<state>,
DDJ(<format>),<result>,<state>,
DCD,<result>,<state>,
ISI(<format>),<result>,<state>,
Transitions,<number_of_transitions>,<transitions_state>,
Scope RJ(<format>),<result>,<state>,
DDPWS,<result>,<state>,
ABUJ(<format>),<result>,<state><NL>
<space>
White space (ASCII 32) character.
<format>
The format value tells you something about how the measurement is made.
For instance, TJ(1E- 12) means that the TJ measurement was derived using
a bit error rate of 1E- 12. A format of (rms) means the measurement is a
root- mean- square measurement. A format of (dd) means the measurement
uses a dual- Dirac delta model to derive the measurement. A format of
(pp) means the measurement is a peak- to- peak measurement.
<result>
The measured results for the RJDJ measurements. A value of 9.37
means that the oscilloscope was unable to make the measurement.
<state>
The measurement result state. See
Table 24
for a list of values and
descriptions of the result state value.
N O T E
This command is only available when the N5400A/N5401A Software is installed.
N O T E
Whether some of these values are included or not depends on the setting of
:MEASure:RJDJ:METHod and :MEASure:RJDJ:REPort.
For example, when :MEASure:RJDJ:REPort or :MEASure:RJDJ:METHod is SPECtral, the
BUJ and ABUJ values are not returned, and there are two PJ values (one "rms" and one
"dd").
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...