:CHANnel<N>:PROBe:EXTernal:GAIN
Command
:CHANnel<N>:PROBe:EXTernal:GAIN <gain_factor>[,{RATio | DECibel}]
The :CHANnel<N>:PROBe:EXTernal:GAIN command sets the probe external
scaling gain factor and, optionally, the units for the probe gain factor. The
reference factors that are used for scaling the display are changed with
this command, and affect automatic measurements and trigger levels.
The RATio or DECibel also sets the mode for the probe attenuation and
also determines the units that may be used for a subsequent command.
For example, if you select RATio mode, then the attenuation factor must
be given in ratio gain units. In DECibel mode, you can specify the units
for the argument as "dB".
<N>
An integer, 1- 4.
<gain_factor>
A real number from 0.001 to 10000 for the RATio gain units, or from
- 60 dB to 80 dB for the DECibel gain units.
Example
This example sets the probe external scaling gain factor for channel 1 to
10.
myScope.WriteString ":CHANNEL1:PROBE:EXTERNAL ON"
myScope.WriteString ":CHANNEL1:PROBE:EXTERNAL:GAIN 10,RATIO"
Query
:CHANnel<N>:PROBe:EXTernal:GAIN?
The :CHANnel<N>:PROBe:EXTernal:GAIN? query returns the probe external
gain setting for the selected channel.
Returned Format
[:CHANnel<N>:PROBe:EXTernal:GAIN] <gain_factor><NL>
Example
This example places the external gain value of the probe on the specified
channel in the variable, varGain, then prints the contents of the variable
to the computer's screen.
myScope.WriteString ":CHANNEL1:PROBE:EXTERNAL ON"
myScope.WriteString ":CHANNEL1:PROBE:EXTERNAL:GAIN?"
varGain = myScope.ReadNumber
Debug.Print FormatNumber(varGain, 0)
N O T E
CHANnel<N>:PROBe:EXTernal command must be set to ON before issuing this command
or query or this command will have no effect.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...