![Agilent Technologies Infiniium 90000 Q-Series Programmer'S Manual Download Page 1002](http://html1.mh-extra.com/html/agilent-technologies/infiniium-90000-q-series/infiniium-90000-q-series_programmers-manual_28686741002.webp)
1002
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
36
Obsolete and Discontinued Commands
:MEASure:CLOCk:METHod
(deprecated)
Command
:MEASure:CLOCk:METHod {FOPLL,<data_rate>,<loop_bandwidth>}
| {EQFOPLL,<data_rate>,<loop_bandwidth>}
| {SOPLL,<data_rate>,<loop_bandwidth>, <damping_factor>}
| {EQSOPLL,<data_rate>,<loop_bandwidth>, <damping_factor>}
| {PCIE,{DEEMphasis | TRANsition | BOTH}}
| {FC,{FC1063 | FC2125 | FC425}}
| {EXPFOPLL,<source>,{RISing | FALLing | BOTH},
<multiplier>,<clock_freq>,<loop_bandwidth>}
| {EXPSOPLL,<source>,{RISing | FALLing | BOTH},
<multiplier>,<clock_freq>,<loop_bandwidth>,<damping_fact>}
| {EXPlicit,<source>,{RISing | FALLing | BOTH}[,<multiplier>]}
| {FIXed,{AUTO | {SEMI[,<data_rate>]} | <data_rate>}}
| {FLEXR,<baud_rate>}
| {FLEXT,<baud_rate>}
The :MEASure:CLOCk:METHod command sets the clock recovery method
to:
•
FOPLL (first order phase- locked loop).
•
SOPLL (second order phase- locked loop).
•
EQFOPLL (equalized first order phase- locked loop).
•
EQSOPLL (equalized second order phase- locked loop).
•
PCIE (PCI Express).
•
FC (Fibre Channel).
•
EXPFOPLL (Explicit First Order PLL).
•
EXPSOPLL (Explicit Second Order PLL).
•
EXPlicit (Explicit Clock).
•
FIXed (Constant Frequency).
•
FLEXR (FlexRay Receiver).
•
FLEXT (FlexRay Transmitter).
N O T E
Some of these command options have been deprecated — options for specifying clock
recovery PLL options have been moved to the new commands
":MEASure:CLOCk:METHod:JTF"
on page 502 and
":MEASure:CLOCk:METHod:OJTF"
on
page 504. See also
":MEASure:CLOCk:METHod"
on page 496.
N O T E
:MEASure:CLOCk commands are available when clock recovery is used by licensed
software (like the E2688A High Speed Serial software or the N5400A/N5401A EZJIT Plus
jitter analysis software).
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...