388
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
20
InfiniiScan (ISCan) Commands
:ISCan:MEASurement:ULIMit
Command
:ISCan:MEASurement:ULIMit <upper_value>
The :ISCan:MEASurement:ULIMit (upper limit) command sets the upper
test limit for the active measurement currently selected by the
:ISCan:MEASurement command.
<upper_value>
A real number.
Example
The following example sets the upper limit of the currently selected
measurement to 500 mV.
myScope.WriteString ":ISCAN:MEASUREMENT:ULIMIT 500E-3"
Suppose you are measuring the maximum voltage of a signal with Vmax,
and that voltage should not exceed 500 mV. You can use the above
program and set the :ISCan:MEASurement:FAIL OUTside command to
specify that the oscilloscope will trigger when the voltage exceeds 500 mV.
Query
:ISCan:MEASurement:ULIMit?
The query returns the current upper limit of the limit test.
Returned Format
[:ISCan:MEASurement:ULIMit] <upper_value><NL>
Example
The following example returns the current upper limit of the limit test and
prints the result to the controller's screen.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString ":ISCAN:MEASUREMENT:ULIMit?"
varValue = myScope.ReadNumber
Debug.Print FormatNumber(varValue, 0)
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...