Calibration Commands
:CALibrate:OUTPut
187
:CALibrate:SKEW
188
:CALibrate:STATus?
189
This chapter briefly explains the calibration of the oscilloscope. It is
intended to give you and the calibration lab personnel an understanding of
the calibration procedure and how the calibration subsystem is intended to
be used.
The commands in the CALibration subsystem allow you to change the
output of the front- panel Aux Out connector, adjust the skew of channels,
and check the status of calibration.
These CALibration commands and queries are implemented in the
Infiniium oscilloscopes:
This chapter briefly explains the calibration of the oscilloscope. It is
intended to give you and the calibration lab personnel an understanding of
the calibration procedure and how the calibration subsystem is intended to
be used.
Oscilloscope
Calibration
Oscilloscope calibration establishes calibration factors for the oscilloscope.
These factors are stored on the oscilloscope's hard disk.
•
Initiate the calibration from the "Utilities Calibration" menu.
You should calibrate the oscilloscope periodically (at least annually), or if
the ambient temperature since the last calibration has changed more than
±5
°
C. The temperature change since the last calibration is shown on the
calibration status screen which is found under the "Utilities Calibration"
dialog. It is the line labeled "Calibration
Δ
Temp: _
°
C."
See also the oscilloscope's
Service Guide
has more details about the
calibration.
Probe Calibration
Probe calibration establishes the gain and offset of a probe that is
connected to a channel of the oscilloscope, and applies these factors to the
calibration of that channel.
•
Initiate probe calibration from the "Setup > Channel > Probes >
Calibrate Probe" menu.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...