:WAVeform:PREamble?
Query
:WAVeform:PREamble?
The :WAVeform:PREamble? query outputs a waveform preamble to the
computer from the waveform source, which can be a waveform memory or
channel buffer.
Returned Format
[:WAVeform:PREamble] <preamble_data><NL>
The preamble can be used to translate raw data into time and voltage
values. The following lists the elements in the preamble.
<preamble_
data>
<format>, <type>, <points>, <count> , <X increment>, <X origin>, < X
reference>, <Y increment>, <Y origin>, <Y reference>, <coupling>, <X display
range>, <X display origin>, <Y display range>, <Y display origin>, <date>,
<time>, <frame model #>, <acquisition mode>, <completion>, <X units>, <Y
units>, <max bandwidth limit>, <min bandwidth limit>
<format>
0 for ASCii format. 1 for BYTE format. 2 for WORD format. 3 for LONG
format. 4 for LONGLONG
<type>
Returned type values can be:
•
1 — RAW
•
2 — AVERage
•
3 — VHIStogram
•
4 — HHIStogram
•
5 — not used
•
6 — INTerpolate
•
7 — not used
•
8 — not used
•
9 — DIGITAL
•
10 — PDETect
<points>
The number of data points or data pairs contained in the waveform data.
(See
":ACQuire:POINts[:ANALog]"
on page 159.)
<count>
For the AVERAGE waveform type, the count value is the fewest number of
hits for all time buckets. This value may be less than or equal to the value
requested with the :ACQuire:AVERage:COUNt command. For RAW and
INTerpolate waveform types, this value is 0 or 1. The count value is
ignored when it is sent to the oscilloscope in the preamble. (See
":WAVeform:TYPE?"
on page 976 and
":ACQuire[:AVERage]:COUNt"
on
page 149.)
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...