:TRIGger:SEQuence:RESet:ENABle
Command
:TRIGger:SEQuence:RESet:ENABle {{ON | 1} | {OFF | 0}}
This command turns the Reset feature on or off for the sequential trigger.
The Reset feature allows you to specify a length of time such that if this
time is exceeded between when the TERM1 event occurs and when the
TERM2 event occurs, the sequential trigger is reset and the oscilloscope
returns to looking for the TERM1 event without triggering. If the Delay
feature (remote command :WAIT) is used as well then the Reset timer does
not start counting down until after the delay period is complete.
You can also base the Reset condition on a logical pattern. If the specified
pattern is found between when the TERM1 occurs and the TERM2 event
occurs, the sequential trigger resets and goes back to looking for the
TERM1 event without triggering. The delay feature does not impact a
logical pattern Reset as the pattern is searched for immediately after the
TERM1 event occurs regardless of whether of not the Delay period is
complete.
If the Reset feature is enabled, the AND qualifier cannot be used for the
TERM1 state.
Query
:TRIGger:SEQuence:RESet:ENABle?
The query returns whether or not the Reset feature is enabled.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...