Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
25
Measure Commands
:MEASure:HISTogram:M2S
Command
:MEASure:HISTogram:M2S [<source>]
The :MEASure:HISTogram:M2S command places the histogram percentage
of points within two standard deviations of the mean measurement into
the Measurements tab of the oscilloscope's user interface.
The source is specified with the :MEASure:SOURce command or with the
optional parameter following the command.
The :MEASure:HISTogram commands only apply to Meas Histogram math
functions, the histogram waveform, or memories containing histograms.
<source>
{ FUNCtion<F> | WMEMory<N> | HISTogram}
<F>
An integer, 1- 16.
<N>
An integer, 1- 4.
Example
This example measures the percentage of points that are within two
standard deviations of the mean of the histogram whose source is
specified using the MEASure:SOURce command.
myScope.WriteString ":MEASure:HISTogram:M2S WMEMory1"
Query
:MEASure:HISTogram:M2S? [<source>]
The :MEASure:HISTogram:M2S? query returns the measurement of the
percentage of points within two standard deviations of the mean of the
histogram.
Returned Format
[:MEASure:HISTogram:M2S]<value>[,<result_state>]<NL>
<value>
The percentage of points within two standard deviations of the mean of
the histogram.
<result_state>
If SENDVALID is ON, the result state is returned with the measurement
result. Refer to the MEASure:RESults command, for a list of the result
states.
Example
This example returns the percentage of points within two standard
deviations of the mean of the current histogram and prints the result to
the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF"
' Response headers off.
myScope.WriteString ":MEASure:HISTogram:M2S? WMEMory1"
varHistm2s = myScope.ReadNumber
Debug.Print FormatNumber(varHistm2s, 0)
See Also
•
":FUNCtion<F>:MHIStogram"
on page 350
•
":HISTogram:MODE"
on page 374
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...