:CHANnel<N>:PROBe
Command
:CHANnel<N>:PROBe <attenuation_factor>[,{RATio | DECibel}]
The :CHANnel<N>:PROBe command sets the probe attenuation factor and
the units (ratio or decibels) for the probe attenuation factor for a
user- defined probe.
The DECibel and RATio parameters also set the "mode" for the probe
attenuation. These parameters, along with attenuation factor, determine
the scaling of the display and affect automatic measurements and trigger
levels.
This mode also determines the units (ratio or decibels) that may be used
for a subsequent command.
<N>
An integer, 1- 4
<attenuation
_factor>
A real number from 0.0001 to 1000 for the RATio attenuation units or
from - 80 dB to 60 dB for the DECibel attenuation units.
Example
This example sets the probe attenuation factor for a 10:1 probe on channel
1 in ratio units.
myScope.WriteString ":CHANNEL1:PROBE 10,RAT"
Query
:CHANnel<N>:PROBe?
The :CHANnel<N>:PROBe? query returns the current probe attenuation
setting and units for the selected channel.
Returned Format
[:CHANnel<N>:PROBe] <attenuation>,{RATio | DECibel}<NL>
Example
This example places the current attenuation setting for channel 1 in the
string variable, strAtten, then the program prints the contents.
Dim strAtten As String
' Dimension variable.
myScope.WriteString ":CHANNEL1:PROBE?"
strAtten = myScope.ReadString
Debug.Print strAtten
If you use a string variable, the query returns the attenuation value and
the factor (decibel or ratio). If you use an integer variable, the query
returns the attenuation value. You must then read the attenuation units
into a string variable.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...