Digital Commands
14
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
273
:DIGital<N>:THReshold
Command
:DIGital<N>:THReshold {CMOS50 | CMOS33 | CMOS25 | ECL | PECL | TTL
| DIFFerential | <value>}
The :DIGital<N>:THReshold command sets the logic threshold value for a
pod. Setting the threshold for digital channels 0 through 7 sets the
threshold for pod 1 while setting the threshold for digital channels 8
through 15 sets the threshold for pod 2. This command is equivalent to
the POD<N>:THReshold command.
The threshold is used for triggering purposes and for displaying the digital
data as high (above the threshold) or low (below the threshold). The
voltage values for the predefined thresholds are:
•
CMOS50 = 2.5 V
•
CMOS33 = 1.65 V
•
CMOS25 = 1.25 V
•
ECL = - 1.3 V
•
PECL = 3.7 V
•
TTL = 1.4 V
•
DIFFerential = 0 V
<N>
An integer, 0- 15.
<value>
A real number representing the voltage value which distinguishes a 1 logic
level from a 0 logic level. Waveform voltages greater than the threshold are
1 logic levels while waveform voltages less than the threshold are 0 logic
levels.
The range of the threshold voltage is from - 3.75 volts to 3.75 volts.
Example
This example sets the threshold to 1.8 volts for bits D15 through D8.
myScope.WriteString ":DIGital8:THReshold 1.8"
Query
:DIGital<N>:THREShold?
The :DIGital<N>:THReshold? query returns the threshold value for the
specified pod.
Returned Format
[:DIGital<N>:THReshold] {CMOS50 | CMOS33 | CMOS25 | ECL | PECL | TTL
| DIFF | <value>}<NL>
N O T E
The DIGital commands only apply to the MSO oscilloscopes.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...