680
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
25
Measure Commands
:MEASure:VMIN
Command
:MEASure:VMIN [<source>]
The :MEASure:VMIN command measures the absolute minimum voltage
present on the selected source waveform. Sources are specified with
:MEASure:SOURce or with the optional parameter following the
:MEASure:VMIN command.
<source>
{CHANnel<N> | FUNCtion<F> | WMEMory<N> | CLOCk | MTRend |
MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source.
<N>
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
FUNCtion<F> is an integer, 1- 16.
Example
This example measures the absolute minimum voltage on the waveform.
myScope.WriteString ":MEASURE:VMIN CHANNEL1"
Query
:MEASure:VMIN? [<source>]
The :MEASure:VMIN? query returns the measured absolute minimum
voltage or minimum FFT amplitude present on the selected source
waveform.
Returned Format
[:MEASure:VMIN] <value>[,<result_state>]<NL>
<value>
Absolute minimum voltage present on the waveform.
<result_state>
If SENDvalid is ON, the result state is returned with the measurement
result. See the :MEASure:RESults table in this chapter for a list of the
result states.
Example
This example returns the measured absolute minimum voltage on the
waveform to the numeric variable, varMinimum, then prints the contents
of the variable to the computer's screen.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString ":MEASURE:VMIN? CHANNEL1"
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...