Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
6
Status Reporting
Mask Test Event Enable Register
For any of the Mask Test Event Register bits to generate a summary bit,
you must first enable the bit. Use the MTEE (Mask Test Event Enable)
command to set the corresponding bit in the Mask Test Event Enable
Register. Set bits are read with the MTEE? query.
Example
Suppose your application requires an interrupt whenever a Mask Test Fail
occurs in the mask test register. You can enable this bit to generate the
summary bit by sending:
myScope.WriteString ":MTEE " + CStr(CInt("&H2"))
Whenever an error occurs, the oscilloscope sets the MASK bit in the
Operation Status Register. Because the bits in the Operation Status Enable
Register are all enabled, a summary bit is generated to set bit 7 (OPER) in
the Status Byte Register.
If bit 7 (OPER) in the Status Byte Register is enabled (via the *SRE
command), a service request interrupt (SRQ) is sent to the external
computer.
N O T E
Disabled Mask Test Event Register Bits Respond, but Do Not Generate a Summary
Bit
Mask Test Event Register bits that are not enabled still respond to their corresponding
conditions (that is, they are set if the corresponding event occurs). However, because they
are not enabled, they do not generate a summary bit in the Operation Status Register.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...