Measure Commands
25
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
489
:MEASure:CGRade:EHEight
Command
:MEASure:CGRade:EHEight <algorithm>
The :MEASure:CGRade:EHEight command enables the eye height
measurement on the current eye pattern. Before using this command or
query, you must use the :DISPlay:CGRade command to enable the color
grade persistence feature. Also, there must be a full eye diagram on screen
before a valid measurement can be made.
<algorithm>
{MEASured | EXTRapolated} EXTRapolated is optional because it is the
default if you do not specify an algorithm.
MEASured will measure the eye height within the window (see
CGRade:EWINdow) of the current data. The smallest eye height is
reported. Extrapolated will estimate the eye height based upon the mean
and standard deviation of the eye top and base.
Example
This example enables the eye height measurement.
myScope.WriteString ":MEASURE:CGRADE:EHEIGHT"
Query
:MEASure:CGRade:EHEight?
The :MEASure:CGRade:EHEight? query returns the eye height measurement
of the color grade display. Before using this command or query, you must
use the :DISPlay:CGRade command to enable the color grade persistence
feature.
Returned Format
[:MEASure:CGRade:EHEight]<value>[,<result_state>]<NL>
<value>
The eye height.
<result_state>
If SENDVALID is ON, the result state is returned with the measurement
result. Refer to the MEASure:RESults command, for a list of the result
states.
Example
This example places the current eye height in the numeric variable,
varValue, then prints the contents of the variable to the computer's screen.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString ":MEASURE:CGRADE:EHEIGHT?"
varValue = myScope.ReadNumber
Debug.Print FormatNumber(varValue, 0)
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...