Agilent Infiniium 90000 Series Oscilloscopes
Programmer's Reference
19
Histogram Commands
:HISTogram:AXIS
373
:HISTogram:MODE
374
:HISTogram:SCALe:SIZE
375
:HISTogram:WINDow:DEFault
376
:HISTogram:WINDow:SOURce
377
:HISTogram:WINDow:LLIMit
378
:HISTogram:WINDow:RLIMit
379
:HISTogram:WINDow:BLIMit
380
:HISTogram:WINDow:TLIMit
381
The HISTogram commands and queries control the histogram features. A
histogram is a probability distribution that shows the distribution of
acquired data within a user- definable histogram window.
You can display the histogram either vertically, for voltage measurements,
or horizontally, for timing measurements.
The most common use for histograms is measuring and characterizing
noise or jitter on displayed waveforms. Noise is measured by sizing the
histogram window to a narrow portion of time and observing a vertical
histogram that measures the noise on a waveform. Jitter is measured by
sizing the histogram window to a narrow portion of voltage and observing
a horizontal histogram that measures the jitter on an edge.
Histograms and
the database
The histograms, mask testing, and color grade persistence use a specific
database that uses a different memory area from the waveform record for
each channel. When any of these features are turned on, the oscilloscope
starts building the database. The database is the size of the graticule area.
Behind each pixel is a 21- bit counter that is incremented each time data
from a channel or function hits a pixel. The maximum count (saturation)
for each counter is 2,097,151. You can use the DISPlay:CGRade:LEVels
command to see if any of the counters are close to saturation.
The database continues to build until the oscilloscope stops acquiring data
or all both features (color grade persistence and histograms) are turned
off. You can clear the database by turning off all three features that use
the database.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...