Measure Commands
:MEASure:DELTatime:DEFine
Command
:MEASure:DELTatime:DEFine <start_edge_direction>,<start_edge_number>,
<start_edge_position>,<stop_edge_direction>,<stop_edge_number>,
<stop_edge_position>
The :MEASure:DELTatime:DEFine command sets the type of direction, the
number of the edge, and the edge position for the delta time measurement.
<start_edge
_direction>
{RISing | FALLing | EITHer} for start directions.
<start_edge
_number>
An integer from 1 to 65534 for start edge numbers.
<start_edge
_position>
{UPPer | MIDDle | LOWer} for start edge positions.
<stop_edge
_direction>
{RISing | FALLing | EITHer} for stop directions.
<stop_edge
_number>
An integer from 1 to 65534 for stop edge numbers.
<stop_edge
_position>
{UPPer | MIDDle | LOWer} for stop edge positions.
Example
This example sets the delta time starting edge to a rising edge on the 5th
edge at the middle position and the stopping edge to falling on the 50th
edge at the lower position.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString _
":MEASURE:DELTATIME:DEFINE RISING,5,MIDDLE,FALLING,50,LOWER"
Query
:MEASure:DELTatime:DEFine?
The :MEASure:DELTatime:DEFine? query returns the measured delta time
value.
Returned Format
[:MEASure:DELTatime:DEFine] <start_edge_direction>,<start_edge_number>,
<start_edge_position>,<stop_edge_direction>,<stop_edge_number>,
<stop_edge_position><NL>
Example
This example places the current value of delta time definition in the string
variable, strValue, then prints the contents of the variable to the
computer's screen. This example assumes the source was set using
:MEASure:SOURce.
Dim strValue As String
' Dimension variable.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString ":MEASURE:DELTATIME:DEFINE?"
strValue = myScope.ReadString
Debug.Print strValue
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...