System Commands
31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
813
:SYSTem:LONGform
Command
:SYSTem:LONGform {{ON | 1} | {OFF | 0}}
The :SYSTem:LONGform command specifies the format for query
responses. If the LONGform is set to OFF, command headers and alpha
arguments are sent from the oscilloscope in the short form (abbreviated
spelling). If LONGform is set to ON, the whole word is output.
Example
This example sets the format for query responses from the oscilloscope to
the short form (abbreviated spelling).
myScope.WriteString ":SYSTEM:LONGFORM OFF"
Query
:SYSTem:LONGform?
The :SYSTem:LONGform? query returns the current state of the
:SYSTem:LONGform command.
Returned Format
[:SYSTem:LONGform] {1 | 0}<NL>
Example
This example checks the current format for query responses from the
oscilloscope, and places the result in the string variable, strResult. Then, it
prints the contents of the variable to the computer's screen.
Dim strResult As String
' Dimension variable.
myScope.WriteString ":SYSTEM:LONGFORM?"
strResult = myScope.ReadString
Debug.Print strResult
N O T E
LONGform Does Not Affect Input Headers and Arguments
LONGform has no effect on input headers and arguments sent to the instrument. You may
send headers and arguments to the oscilloscope in either the long form or short form,
regardless of the current state of the :SYSTem:LONGform command.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...