522
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
25
Measure Commands
:MEASure:DELTatime
Command
:MEASure:DELTatime [<source>[,<source>]]
The :MEASure:DELTatime command measures the delta time between two
edges. If one source is specified, the delta time from the leading edge of
the specified source to the trailing edge of the specified source is
measured. If two sources are specified, the delta time from the leading
edge on the first source to the trailing edge on the second source is
measured.
Sources are specified with the :MEASure:SOURce command or with the
optional parameter following the :MEASure:DELTatime command. The rest
of the parameters for this command are specified with the
:MEASure:DEFine command.
The necessary waveform edges must be present on the display. The query
will return 9.37 if the necessary edges are not displayed.
<source>
{CHANnel<N> | FUNCtion<F> | DIGital<M> | WMEMory<N> | CLOCk |
MTRend | MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source.
<N>
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
FUNCtion<F> is an integer, 1- 16.
<M>
An integer, 0- 15. Digital channels are available on mixed- signal
oscilloscopes.
Example
This example measures the delta time between channel 1 and channel 2.
myScope.WriteString ":MEASure:DELTatime CHANnel1,CHANnel2"
Query
:MEASure:DELTatime? [<source>[,<source>]]
The :MEASure:DELTatime? query returns the measured delta time value.
Returned Format
[:MEASure:DELTatime] <value>[,<result_state>]<NL>
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...