Measure Commands
25
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
599
:MEASure:PBASe
Command
:MEASure:PBASe <source>, <pulse width percent>
The :MEASure:PBASe command measures the average of the data of a
negative pulse within the pulse window. The pulse window is a range of
data centered within the pulse width using the specified percentage of the
data as measured as the middle threshold level. For example, a 50%
window would not include in the average the first or last 25% of the pulse
width as measured at the middle threshold level. A 100% window would
measure the average of the entire positive or negative pulse. In measure
all edges mode and EZJIT, these measurements can be trended,
histogrammed, etc.
<source>
{CHANnel<N> | FUNCtion<F> | WMEMory<N>}
<N>
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
FUNCtion<F> is an integer, 1- 16.
<pulse width
percent>
pulse width percent to use in average (in percent, 0- 100)
Example
This example measures the average of the data of a negative pulse within
the pulse window (width set to 50%)
myScope.WriteString ":MEASURE:PBASe CHAN1, 50"
Query
:MEASure:PBASe? <source>, <pulse width percentage>
The :MEASure:PBASe? query returns the average pulse base of the data of
a negative pulse within the specified pulse window.
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...