Acquire Commands
9
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
159
:ACQuire:POINts[:ANALog]
(Memory depth)
Command
:ACQuire:POINts[:ANALog] {AUTO | <points_value>}
The :ACQuire:POINts[:ANALog] command sets the requested analog
memory depth for an acquisition. Before you download data from the
oscilloscope to your computer, always query the points value with the
:WAVeform:POINts? query or :WAVeform:PREamble? query to determine the
actual number of acquired points.
You can set the points value to AUTO, which allows the oscilloscope to
select the optimum memory depth and display update rate.
<points_value>
An integer representing the memory depth.
The range of points available for a channel depends on the oscilloscope
settings of sampling mode, sampling rate, and trigger sweep. The following
tables show the range of memory values for the different memory options.
Table 6
1G Memory Option Installed
Sampling mode and sample rate
Trigger Sweep
Single
Auto or Triggered
Normal, Peak Detect (full sample rate),
and High Resolution (full sample rate)
Modes
16 to 1,049,600,000
Normal and High Resolution modes
with averaging
16 to 2,050,000
High Resolution Mode at < full sample
rate
16 to 524,800,000
16 to 262,400,000
Peak Detect Mode at < full sample
rate
16 to 262,400,000
16 to 131,200,000
Equivalent Time Mode
16 to 262,144
Table 7
500M Memory Option Installed
Sampling mode and sample rate
Trigger Sweep
Single
Auto or Triggered
Normal, Peak Detect (full sample rate),
and High Resolution modes
16 to 512,500,000
Normal and High Resolution modes
with averaging
16 to 2,050,000
Summary of Contents for Infiniium 90000 Q-Series
Page 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Page 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Page 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Page 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Page 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Page 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Page 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Page 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Page 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Page 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Page 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Page 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Page 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Page 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Page 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Page 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Page 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Page 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...