9.5.2. Setup the Input Channel
The settings differ according to whether Sensor is set to the thermocouple (TC) or platinum
resistance temperature detector (RTD) type.
<TC>
<RTD>
(1)
(2)
(1)
(2)
(3)
(4)
(3)
(4)
(5)
(6)
(1) Sensor:
Select [TC] or [RTD] as the sensor type.
(2) Type:
For a thermocouple (TC) sensor, select K, E, J, T, N, R, S, B, or C as the thermocouple
(TC) to connect.
For RTD, select Pt100 (1 mA), Pt100 (0.5 mA), or Pt1000 (0.1 mA).
(3) Meas. range: Set the temperature range to measure. Three types of ranges can be selected for each
sensor type.
(4) Refresh data: Select low speed (1 s), medium speed (100 ms), or high speed (1.5 ms) as the data
update speed.
(5) RJC:
Select [Internal] to directly connect the sensor when using a thermocouple (TC).
When placing the reference junction (cooling point) outside, select [External] to perform
temperature compensation.
(6) LFD:
Set [ON] or [OFF]. When set to [ON], the output goes over the scale when there is a
disconnection.
The Meas. , Color , Disp. pos. , Disp. range , Disp. min , Disp. max , Sheet , available
setting ranges, and measurement value settings are the same as for other modules.